508 Interface Module User’s Guide
508 Interface Module User’s Guide World Headquarters Gilson, Inc. 3000 Parmenter Street P.O. Box 620027 Middleton, WI 53562-0027 USA Telephone: 608-836-1551 Fax: 608-831-4451 Gilson S.A.S. 19, avenue des Entrepreneurs F-95400 VILLIERS LE BEL France www.gilson.com sales@gilson.com, service@gilson.com, training@gilson.com ©2006 Gilson, Inc. All Rights Reserved.
Table of Contents 1 Introduction Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2 Standard Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2 Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2 Customer Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3 Technical Specifications . . . . . . . . . . . . .
Introduction 1 The Gilson 508 Interface Module connects your computer to many instruments made by Gilson. It permits bi-directional communication by interconverting the RS-232 signal levels used by computers and the RS-422/485 signal levels required by the Gilson Serial Input/Output Channel (GSIOC).
1 Unpacking Introduction Unpacking Unpack the interface module and its accessories carefully from the carton. Verify that all parts are included and are undamaged. Do this now, even if the interface module will not be used immediately. Many carriers must receive concealed damaged claims within seven days of delivery. Please retain all packing material so the unit may be shipped safely, if necessary.
Introduction 1 Customer Service Customer Service Gilson, Inc. and its worldwide network of authorized representatives provide customers with the following types of assistance: sales, technical support, applications, and instrument repair. If you need assistance, please contact your Gilson-authorized representative. Specific contact information can be found on the Gilson website at www.gilson.com. To help us serve you quickly and efficiently, please refer to Repair and Return Policies on page 5-3.
1 Technical Specifications Introduction Technical Specifications Please be aware of the following before connecting the interface module. Warning! Changes or modifications to the interface module not expressly approved by Gilson could void the factory-authorized warranty. The interface module has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC commercial environment.
Installation 2 The 508 Interface Module is pre-configured as an RS-232-to-GSIOC interface. Turn off and unplug all computers and instruments before making the connections described in this section.
2 Electrical Connections Installation Electrical Connections Rear Panel 508 Interface Module Rear Panel RS-232 Locate the RS-232 cable provided with the interface module. Attach the male end of the RS-232 cable to the 508 RS-232 input. Tighten the retaining screws. Connect the female end of the RS-232 cable to the computer’s RS-232 serial communications port. Again tighten the retaining screws.
2 Installation Electrical Connections Refer to the diagram below to connect the 508 to another Gilson instrument using a GSIOC cable. 9-pin female connector; connect to the GSIOC port on the interface module 9-pin female connector; connect to GSIOC port of GSIOC device 9-pin male connector; connect to another GSIOC cable for linking additional GSIOC devices With additional GSIOC cables, you can connect as many as 32 devices.
Operation 3 The 508 Interface Module provides RS-232-to-GSIOC interface from the computer. Gilson control software packages let you issue GSIOC commands that communicate with any GSIOC device connected to the interface module.
3 Status Lights Operation Status Lights The status lights on the 508 Interface Module are useful for verifying operation. If the Power light is off, power is not being sent to the interface module. If the RS-232 light is flashing, the 508 Interface Module is responding to the RS-232 input. If the GSIOC light is flashing, the 508 Interface Module is receiving a response from a GSIOC device.
Maintenance 4 The 508 Interface Module should not require maintenance. The best precautions for continued operation are to avoid spilling liquids on the connectors and to avoid mechanical strain on the wire connections.
Troubleshooting 5 When using the 508 Interface Module with Gilson-supplied accessories, computers and application software, troubleshooting is simple. Most problems are due to improper connections and installation.
5 Troubleshooting Chart Troubleshooting Troubleshooting Chart No response to RS-232 input • There is no DC power connection at power input. If the Power light is off, check that transformer is plugged into an outlet. • RS-232 connections are not correct. Check for proper cable and connectors. • Check RS-232 port of computer for proper operation with other RS-232 devices. GSIOC devices do not respond to commands 5-2 • Check GSIOC cable connections.
Troubleshooting 5 Repair and Return Policies Repair and Return Policies Before Calling Us Gilson-authorized representatives will be able to serve you more efficiently if you have the following information: • the serial number and model number of the instruments involved. The serial number is located on the bottom of the unit.
5 Repair and Return Policies Troubleshooting Return Procedure Contact your local distributor’s Customer Service Department to obtain authorization before returning any Gilson equipment. To return a piece of equipment: • Carefully pack the unit to prevent damage in transit. Check with your distributor regarding proper method of shipment. No responsibility is assumed by Gilson or your distributor for damage caused by improperly packaged instruments.