User's Manual

FIELD EQUIPMENT
NUSEIS
Geophysical Technology Inc Field Equipment Page | 47
There are no electronics inside the charging panels, they are simply distribution points.
When using the USB Data Management Panels, it is possible to change between different
configuration options, such as having either 1, 2, or 3 Data Management Panels installed in the
portable case.
o When using the optional TransferJet Data Management Panels it is advisable to only
have 1 panel installed per portable case.
Use of the Portable Data Management Case
Follow the NRU cleaning steps in the Cleaning section .
Charging can be carried out in the 2 charging panels as it is with the Portable Charging Case, see
the previous chapter for details on charging with the Portable Case.
To carry out NRU uploading or commissioning operations, the Portable Case must be connected
to a NuScribe computer via the external ethernet port seen at the bottom of figure 32 below.
Figure 32
Once the NRUs are loaded into the offload ports, follow the NuScribe manual directions for
either offload or commissioning . See NuScribe manual page 63.
The charger and offload ports should be monitored for a buildup of dirt and debris in the
bottom of the cups. Too much dirt or debris may keep the NRU from turning on or uploading
properly due to the debris obstructing proper seating.
o Remove power from the case
o Use compressed air, preferably with a line dryer, to blow out the cups.
o No CRC or other aerosol cleaning sprays should be used to clean the cups. A dry soft
brush can be used to loosen debris in the cup prior to blowing it out with compressed
air.