User Manual
Geophysical Survey Systems, Inc. Model 5100 Antenna
System Settings and User Notes
Manual MN31-108 Rev - 3
• The green button on the cart must be pressed continuously during calibration and operation.
• Releasing the button for less than one second will place a marker in the data.
• During data acquisition, releasing the button for two seconds will place the system in the Standby
mode. In this mode, momentarily releasing the button will restart data collection.
• Releasing the button for longer than 10 seconds will close any open files and turn off the
transmitter. To resume, the data acquisition sequence must be repeated.
A long yellow handle may be added to the cart. This has a muli-function switch at the top end.
• If you are using the long yellow handle attached to the StructureScan cart, you will need to keep
the red “kill” switch on the handle depressed at all times.
• Releasing the handle for 2 - 10 seconds pauses data collection. Data collection that is in Pause
mode can be restarted by briefly releasing the “kill switch” or pressing the marker button.
• If the “kill” switch is released for more than 10 seconds, the file will automatically be closed.
• Markers are added to the data by pressing the marker button on the end of the handle, or by
quickly releasing and depressing the kill switch.
Red “kill” switch. Must
be depressed during
calibration and data
collection.
Marker button.
Press to add
markers to data.
Handle of yellow
extension arm to
cart.