Use fine tuning to set
the vertical angle as
V1+3’, focus on the
reticle center accurately,
then turn the instrument
counterclockwise 90 °,
wait for stable value,
press [F4](OK).
[F4]
ƻ
6 After finishing, it
will display the results,
press [F4](OK), save and
back to menu.
[F4]
Note: CoK (linear coefficient): If absolute value > 1.5, you need to re-calibrate;
In the correction process by pressing the ESC key, will exit, holding compensator
parameters unchanged.
1.5 Instrument constant setting
In Tools Menu, choose “1 Adjust”, and then press [F4 to enter “Const. Setting”.
Press [F4](OK) to save after editing the constants.
1.6 Factory setting
In Tools Menu, choose “1 Adjust”, and then press [5] to enter “Factory Setting”.
If you need to reset the instrument parameters to factory state, you can use this
function, press key [F4] (Yes) and then the instrument will auto power off.