Note: CoK (linear coefficient): If absolute value > 1.5, you need to re-calibrate;
In the correction process by pressing the ESC key, will exit, holding compensator
parameters unchanged.
1.4 Adjust Tilt Y
In Tools Menu, choose “1 Adjust”, and then press [F4] to enter “Adjust Tilt Y”.
These are the calibaration of y-diretion of compensator’s vertical axis.
Steps Key Display
ƻ
1 Level instrument,
focus on the reticle of
collimator, record the
vertical angle V0. Use
fine tuning to set
vertical angle to
V0+3’,focus on the
reticle center accurately,
then turn the instrument
counterclockwise 90 °,
wait for stable
[F4]
V1-3’,focus on the
reticle center accurately,
wait for stable value,
pressF4(OK).
ƻ
5 Use fine tuning to set
the vertical angle as
V1+3’, focus on the
reticle center accurately,
wait for stable value,
press [F4](OK).
[F4]
ƻ
6 After finishing, it
will display the results,
press [F4](OK), save and
back to menu.
[F4]