User Manual
Total Station Manual
191
1.2 Adjust Index Error
In Tools Menu, choose “1 Adjust”, then press [F2] to enter “Adjust Index Error”.
6
6WHSV
Steps Key Display
ƻ
1
After leveling the
total station, aim at
target with face left,
then press [F4](OK).
[F4]
ƻ
2 Aim at the same
target with face right,
and press [F4] (OK).
[F4]
ƻ
3
Program will show
the result value, press
[F4](OK) to save.
[F4]
Note: If there is no special requirement, the compensator should be turned on before
Index error correction.
1.3 Adjust Tilt X
Before compensating for the compensator, make sure that the indicator
difference is recalibrated in accordance with 9.1.2 procedure in the closed
compensator state.
First, place the instrument as picture shown below with collimator facing up.
This will help screw A to adjust the inclination of the instrument.