Tribrach: If the bubble extends beyond the circle, adjust it using the adjustment pin in conjunc-
tion with the adjustment screws. Turn the adjustment screws:
•
To the left: and the bubble approaches the screw.
•
To the right: and the bubble goes away from the screw.
4 Repeat step 3 on the instrument and tribrach until both levels are centered and no further
adjustments are necessary.
After the calibration, no adjustment screw should be loose.
6.1.5 Inspecting the Laser Plummet of the Instrument
The laser plummet is integrated into the vertical axis of the instrument. Under normal conditions
of use, the laser plummet does not need adjusting. If an adjustment is necessary due to external
influences, the instrument has to be returned to a GeoMax service department.
003860_001
1 Set up the instrument on the tripod approximately 1.5 m above the ground and level up.
2 To activate the laser plummet, turn on the instrument, and, if tilt correction is set on, the
laser plummet activates automatically, and the Level up screen appears. Otherwise, tap on
the Level/Face icon in status bar.
☞
Inspection of the laser plummet should be carried out on a bright, smooth and hori-
zontal surface, such as a sheet of paper.
3 Mark the center of the red laser dot on the ground.
4 Turn the instrument slowly through 360°, carefully observing the movement of the red
laser dot.
☞
☞
Inspect the laser plummet
step-by-step
Tools 41