A longitudinal component in direction of the telescope and a transversal component perpendicu-
lar to the telescope define the plane of the dual axis compensator of the instrument.
The longitudinal compensator index error (l) has a similar effect as the vertical index error and
effects all vertical angle readings.
The transversal compensator index error (t) is similar to the tilting axis error. The effect of this
error to the horizontal angle readings is 0 at the horizon and increases with steep sightings.
The vertical index error and the compensator index error are determined simultaneously.
1. Select Tools from the MAIN MENU.
2. Select Calibr. from the TOOLS menu.
3. Select a calibration option from the CALIBRATION screen.
In the CALIBRATION screen there are several calibration options.
Menu selection Description
HA-Collimation Refer to "6.1.3 Calibrating Line-of-Sight, Vertical Index
Error and Compensator Index Error".
Vertical Index Refer to "6.1.3 Calibrating Line-of-Sight, Vertical Index
Error and Compensator Index Error".
View Adjustment Data Displays the current calibration values that have been set
for HA-Collimation and V-index.
The procedures and conditions required to correct line-of-sight and vertical index errors are the
same, therefore the procedure will only be described once.
6.1.4 Calibrating the Level of the Instrument and Tribrach
1 Place and secure the tribrach onto the tripod, and then secure the instrument onto the tri-
brach.
2 Using the tribrach footscrews, level the instrument with the electronic level. To activate the
electronic level, turn on the instrument, and, if tilt correction is set on, the Level up screen
appears automatically. Alternatively, tap on the Level/Face icon in status bar.
3 The bubbles of the instrument and tribrach levels must be centered. If one or both levels
are not centered, adjust as follows.
Instrument: If the bubble extends beyond the lines, use the Allen key supplied to center it with
the adjustment screws.
☞
Access
Calibration options
☞
Calibrate the level step-
by-step
40 Tools