Product Manual

Table Of Contents
ATMOS
To enter atmospheric data ppm.
PPM
To enter an individual ppm value.
SCALE
To enter projection scale details.
FREQ.
To view the EDM frequency.
Field Description
Mode IR-Default Fine measuring mode for high precision measurements
with prisms.
IR-Quick Quick measuring mode with prisms, with higher meas-
uring speed and reduced accuracy.
IR-Continuous For continuous distance measurements with prisms.
Foil For distance measurements using Retro reflective tar-
gets.
RL-Default For distance measurements without prisms.
RL-Continuous For continuous distance measurements without prisms.
Type Circular ZPR100 Abs. Const.: -34.4mm
Custom The user can define their own prism.
Constants can be entered in mm in Abs. Const.
Mini ZMP100 Abs. Const.: -16.9mm
JPMini ZPM100 Abs. Const.: 0.0mm
360° ZPR1 Abs. Const.: -11.3mm
360°Mini GRZ101 Abs. Const.: -4.4mm
Foil ZTM100 Abs. Const.: 0.0mm
None NP-modes Abs. Const.: 0.0mm
Abs.
Const.
This field displays the GeoMax prism constant for the selected Type.
When Type is Custom this field becomes editable to set a user defined con-
stant. Input can only be made in mm.
Limit value: -999.9mm to +999.9mm.
Laser-
Beam
Off Visible laser beam is deactivated.
On Visible laser beam for visualising the target point is
activated.
This screen enables the entry of atmospheric parameters. Distance measurement is influenced
directly by the atmospheric conditions of the air in which the measurements are taken. In order
to take these influences into consideration distance measurements are corrected using atmos-
pheric correction parameters.
The refraction correction is taken into account in the calculation of the height differences and the
horizontal distance. Refer to "9.6 Scale Correction" for the application of the values entered in
this screen.
EDM SETTINGS
ATMOSPHERIC DATA
ENTRY
34 Settings