Product Manual
Table Of Contents
- 1 Safety Directions
- 2 Description of the System
- 3 User Interface
- 4 Operation
- 4.1 Main Menu
- 4.2 System Information
- 4.3 Setting Up the TPS Instrument
- 4.4 Setting up for Remote Control (with the RadioHandle)
- 4.5 Connecting to a Personal Computer
- 4.6 Power Functions
- 4.7 Batteries
- 4.8 Working with the Memory Device
- 4.9 Working with Bluetooth
- 4.10 LED Indicators
- 4.11 Guidelines for Correct Results
- 5 Settings
- 6 Apps
- 7 Care and Transport
- 8 Technical Data
- 8.1 Angle Measurement
- 8.2 Distance Measurement with Reflectors
- 8.3 Distance Measurement without Reflectors
- 8.4 Distance Measurement - Long Range (LO mode)
- 8.5 Automatic Prism Aiming AiM
- 8.6 Prism Search (Scout) - only available at Zoom90
- 8.7 Conformity to National Regulations
- 8.8 General Technical Data of the Product
- 8.9 Scale Correction
- 8.10 Reduction Formulas
- 9 Software Licence Agreement
- 10 Glossary
- Appendix A Menu Tree
- Appendix B Directory Structure
- Appendix C Pin Assignments
8.5 Automatic Prism Aiming AiM
Reflector Range AiM mode TRack mode
[m] [ft] [m] [ft]
Standard prism (ZPR100) 1000 3300 800 2600
360° prism (ZPR1, GRZ122) 800 2600 600 2000
360° Mini prism (GRZ101) 350 1150 200 660
Mini prism (ZMP100) 500 1600 400 1300
Mini prism (ZMP101) 500 1600 400 1300
Reflector tape
60 mm x 60 mm (ZTM100)
45 150 not qualified
☞
The maximum range can be restricted by poorer conditions, for example rain.
Shortest measuring distance: 360° prism AiM: 1.5 m
Shortest measuring distance: 360° prism TRack: 5 m
AiM angle accuracy Hz, V (std. dev. ISO 17123-3): 1 " (0.3 mgon)
Base Positioning accuracy (std.dev.): ± 1 mm
Several factors can influence the system's accuracy for determining the location of a prism:
•
Internal AiM accuracy
•
Angular accuracy of the instrument
•
Type and centring accuracy of the prism
•
Selected EDM measuring program
•
External measuring conditions
Therefore, the overall pointing accuracy of the determined point location can be lower than the
given angular accuracy and the AiM accuracy.
The following paragraphs provide a short overview of these influencing factors and their possible
intensities.
Angular accuracy
The accuracy of angular measurements depends on the instrument type. The angular accuracy
for total stations is typically in the range from 0.5² to 5². The resulting error depends on the
measurement distance.
Angular accuracy
Possible deviation
*
at 100 m distance
1² ~0.5 mm
3² ~1.5 mm
*
Orthogonal to the line of sight.
☞
Refer to the data sheet of the respective instrument model for information about the
angular accuracy.
EDM accuracy
The distance measurement accuracy consists of two parts: a fixed value and a distance-depend-
ent value (ppm-value).
Example: "Single measurements: 1 mm + 1.5 ppm"
The EDM accuracies for prism and reflectorless measurements can differ. Additionally, the accu-
racies can differ depending on the used technologies.
☞
Refer to the appropriate data sheet for information about the EDM accuracy.
Range AiM/TRack
AiM accuracy with the
ZPR100 prism
System Accuracy
Technical Data 55