3. Press REC to measure to the target point.
4. Change face and aim at the target point again
180°
180°
Zoom_014
For checking the horizontal aim, the difference in
HA and VA are displayed.
5. Press REC to measure to the target point.
The old and new calculated values are displayed.
6. Either:
Press OK to save the new calibration data, or
Press ESC to exit without saving the new cal-
ibration data.
The following are important messages or warnings that may appear.
Messages Description
VA-angle not suitable for adjust-
ment !
The vertical angle deviates from the required hori-
zontal / line-of-sight, or in face II the vertical angle
deviates by more than 5° from the target point. Aim
at the target point with an accuracy of min. 5°.
Confirmation of the message required.
Results out of tolerance. Previous
values retained !
Computed values out of tolerance. The previous val-
ues are retained and measurements should be
repeated. Confirmation of the message required.
HA-angle not suitable for adjust-
ment !
Horizontal angle in face II deviates by more than 5°
from the target point. Aim on the target point with
an accuracy of min. 5°. Confirmation of the mes-
sage required.
Measurement Error. Try again. Measurement error appears when, for example,
there is an unstable set up. Repeat the process.
Confirmation of the message required.
Time limit exceeded ! Please
repeat adjustment !
Time difference between measurements for results
storage exceeds 15 minutes. Repeat the process.
Confirmation of the message required.
Messages
120 Calibration