Product Manual
Table Of Contents
- 1 Safety Directions
- 2 Description of the System
- 3 User Interface
- 4 Operation
- 5 Settings
- 6 Tools
- 7 Functions
- 8 Coding
- 9 MapView Interactive Display Feature
- 10 Applications - Getting Started
- 11 Applications
- 12 Data Management
- 13 Calibration
- 14 Care and Transport
- 15 Technical Data
- 15.1 Angle Measurement
- 15.2 Distance Measurement with Reflectors
- 15.3 Distancer, Measurements without Reflectors (Reflectorless mode)
- 15.4 Distance Measurement Reflector (Long Range)
- 15.5 Conformity to National Regulations
- 15.6 General Technical Data of the Product
- 15.7 Scale Correction
- 15.8 Reduction Formulas
- 16 Software Licence Agreement
- 17 Glossary
- Appendix A Menu Tree
- Appendix B Directory Structure
- Appendix C Connector PIN Layout
3. Press REC to measure to the target point.
4. Change face and aim at the target point again
180°
180°
Zoom_014
☞
For checking the horizontal aim, the difference in
HA and VA are displayed.
5. Press REC to measure to the target point.
☞
The old and new calculated values are displayed.
6. Either:
•
Press OK to save the new calibration data, or
•
Press ESC to exit without saving the new cal-
ibration data.
The following are important messages or warnings that may appear.
Messages Description
VA-angle not suitable for adjust-
ment !
The vertical angle deviates from the required hori-
zontal / line-of-sight, or in face II the vertical angle
deviates by more than 5° from the target point. Aim
at the target point with an accuracy of min. 5°.
Confirmation of the message required.
Results out of tolerance. Previous
values retained !
Computed values out of tolerance. The previous val-
ues are retained and measurements should be
repeated. Confirmation of the message required.
HA-angle not suitable for adjust-
ment !
Horizontal angle in face II deviates by more than 5°
from the target point. Aim on the target point with
an accuracy of min. 5°. Confirmation of the mes-
sage required.
Measurement Error. Try again. Measurement error appears when, for example,
there is an unstable set up. Repeat the process.
Confirmation of the message required.
Time limit exceeded ! Please
repeat adjustment !
Time difference between measurements for results
storage exceeds 15 minutes. Repeat the process.
Confirmation of the message required.
Messages
120 Calibration