6 SDI-12-commands and responses
The OTT SVR 100 communicates using the physical RS-485 interface via the SDI-12 transfer protocol. In this technical
documentation, you will find a detailed description of the SDI-12 commands implemented in the SDI-12 transfer protocol.
Further information on the SDI-12 standard can be found in the document "SDI-12; A Serial-Digital Interface Standard for
Microprocessor-Based Sensors; Version 1.3" (see website "www.sdi-12.org").
All advanced, manufacturer-specific SDI-12 commands on the OTT SVR 100 begin with "O" for OTT. With these commands
i
t is possible to configure the OTT SVR 100, for example using the "SDI-12 transparent mode" on a data logger or with the
OTT USB/SDI-12 interface (accessory).
Conventions for measured value formats
p – Sign (+,–)
b – Figure (before the decimal point)
e – Figure (after the decimal point)
! – Ends a command
6.1 Overview of standard SDI-12 commands
! a! Acknowledgement active
! aI! Send identification
! aAb! Change sensor address
! ?! Query sensor address; factory setting: 0
! aM! Start measurement
! aD0! Send data
! aR0! Continuous measurement
! aMC! Start measurement and request CRC (Cyclic Redundancy Check)
! aC! Start concurrent measurement (simultaneous measurement with multiple sensors on one bus line)
! aCC! Start concurrent measurement and request CRC
! aV! Start system test
Advanced commands (manufacturer specific)
! aOAA<value>! Set internal filter type
aOAA! Read internal filter type
Factory setting: 1 ➝ Floating mean
! aOAB<value>! Set measuring sensitivity
aOAB! Read measuring sensitivity
Factory setting: 14
! aOAC<value>! Set filter length
aOAC! Read filter length
Factory setting: 1 ➝ internal filter deactivated
! aOSD<value>! Set flow direction filter
aOSD Read flow direction filter
Factory setting: 0 ➝ Flow direction filter deactivated; both flow directions are recorded
! aOSU<value>! Set unit for "flow velocity" measured values
aOSU Read unit for "flow velocity" measured values
Factory setting: 0 ➝ m/s
17