User's Manual
LX-80 User Manual
Page 28 of 32
Additional
Data
aD1! .. aD9!
a0<CR><LF>
a0<CRC><CR><LF>
No values are returned for additional
data.
Concurrent
Measurement
Send Data
Additional
Data
aC!/aCC!
aD0!
aD1! .. aD9!
ammm7<CR><LF>
a+f+f+d+d+d+d+f<CR><LF>
a+f+f+d+d+d+d+f<CRC><CR>
<LF>
a0<CR><LF>
a0<CRC><CR><LF>
7 values are ready within mmm
seconds
+f – measured relative level depending
on sensor height
+f – measured distance from sensor to
water
+d – measured temperature inside
device
+d – measured water temperature (on
request only)
+d – measured accelerometer angle of
device in x direction (on request only)
+d – measured accelerometer angle of
device in y direction (on request only)
+f – SNR of latest measurement
No values are returned for additional
data.
Continuous
Measurement
aR0! .. aR9!
a<CR><LF>
a<CRC><CR><LF>
The continuous mode is not supported
Get
measurement
unit for level
aXGLUN!
a+d<CR><LF>
+d – measurement unit for level
0 – m
1 – cm
2 – mm
3 – in
4 – ft
Set
measurement
unit for level
aXGLUN+d!
a+d<CR><LF>
+d – measurement unit for level
0 – mm
1 – cm
2 – m
3 – in
4 – ft
Get minimum
deadzone value
aXGDZ0!
a+f<CR><LF>
+f – deadzone minimum value, sensor
will not report measurement lower
than this