User's Manual

LX-80 User Manual
Page 26 of 32
Fun
Data Addr
Data Length
Data Range
Details
0x06
0x0007
2 bytes
1 65535
SDI-12 ID
0x000B
2 bytes
1 1000
default: 100
Number of frames for
average measurement
0x000C
2 bytes
1 65535
default: 0
Minimum spectrum
amplitude threshold, used
for detecting peaks
0x000D
2 bytes
0 device range
default → 0
Deadzone minimum
0x000E
2 bytes
0 device rang
default → device range
Deadzone maximum
0x000F
2 bytes
0 device range
default → device range
4-20mA minimum value
0x0010
2 bytes
0 device range
default → device range
4-20mA maximum value
0x0011
2 bytes
0 device range
Level measurement offset
0x0012
2 bytes
0 1
IR filter constant





0x0015
2 bytes
2 printable characters
Serial number[0-1]
0x0016
2 bytes
2 printable characters
Serial number[2-3]
0x0017
2 bytes
2 printable characters
Serial number[4-5]
0x001C
2 bytes
0-2
0 no filter
1 IIR filter
2 window filtering
0x0022
2 bytes
Sensor height
Sensor height above the
riverbed
0x0023
2 bytes
Staff gauge
Staff gauge measurement
used to indirectly determine
sensor height;
Sensor height = staff gauge +
current measured distance
from sensor to water