FA50 Operation Manual 61
reagent card with sample and confirm the sample information”,
if you need to change the sample type, click “Cancel”, reselect
the sample type, click “STD Mode” to start the test again; if the
current test item is the same as the previous test item, the
analyzer will enter the “STD Mode” process; if the current test
item is different from the last test item, the instrument will
prompt “Item information changed, please confirm the
information before starting the test”. After confirming that the
item information and sample information are correct, click
“STD Mode” again, and the analyzer will start the countdown.
When the countdown is over, the measurement will start
automatically. After the test is completed, the result will be
displayed in the result display box in the lower right corner of
the measurement interface.
4.If fast measurement is required, please place the sampled
reagent card at room temperature for 15min (to ensure the
accuracy of the result, please use a timer to time), and quickly
insert the reagent card into the analyzer's test slot (sample
hole facing in), press the “Fast Mode” button. When the scan
code setting status of the reagent card is Auto, click “Fast
Mode” and the analyzer will recognize the barcode on the
reagent card and automatically switch to the item to be tested.
At the same time, prompt “Please insert the incubated reagent
card and confirm the sample information”; if you need to
change the sample type, click “Cancel”, and then select the
sample type again, click “Fast Mode” to start the measurement
again. If the current test item is the same as the last test item,
the analyzer will start measurement; if the current test item is