24 FA50 Operation Manual
Fig. 5-5-7
5.5.4 Fast Mode
If the current reagent batch has already been swiped, there is no
need to swipe the card; if you replace the new batch of reagent
cards, when you replace the new batch of reagent cards, you need
to swipe the card again. For details on the card swiping process,
see 5.8.3.6 Swipe card.
When the scan code setting status of the reagent card is automatic,
click Fast Mode, and the analyzer will recognize the barcode on
the reagent card and automatically switch to the item to be tested.
And prompt Please insert the incubated reagent card and confirm
the sample information”, as shown in Fig. 5-5-8; if you need to
change the sample type, click Cancel. After selecting the sample
type again, click Fast Mode to start the measurement again. If the