Installation Instructions

Error! Use the Home tab to apply 标题 1 to the text that you
want to appear here.
FA50 Operation Manual 7
If all the items are there, carefully take out the analyzer and put it on
a stable countertop. Do not place the analyzer near the wall or
other places where are difficult for disconnection, for it is a
non-permanently connected device.
2.3.2 Connecting the power cable
Make sure the power switch at the rear panel is turned off. Plug one
end of the power cable into the instrument’s power port, another
end into the power supply interface.
The specification of the fuse used for this instrument is F3AL250V,
with rated current of 3A and rated voltage of 250V, which is a
double cap fast-blow glass tube lead fuse.
2.3.3 Connecting the equipotential terminal
When need to work together between different devices, in order to
achieve certain functions, a same reference potential is required, at
this time you need to connect the equipotential terminal. Inhibit the
potential difference by equipotential connection to eliminate the
electromagnetic interference.
To connect the equipotential terminal, take out the supplied ground
cable and directly connect the equipotential terminal at the rear of
the instrument to the earth wire through the ground cable.
2.3.4 Installing (replacing) the printing paper
1)Slightly press to open the cover of the built-in printer.