78 FA50 Operation Manual
A.1.3 Stability
After turning on the analyzer for 4 hours and 8 hours, test the
reference material (calibrator) with a fixed value, the bias (B) of its
average value and marked value should not be greater than 10%.
A.1.4 Linearity
The correlation coefficient γ
2
should be equal or greater than 0.990.
A.2 Supporting Reagents
There are corresponding test cards for different items.
A.3 Test Parameter Information
Category
Item
No.
Item Name
Full Name
Unit
Infection
1
hs-CRP
High sensitive
C-reactive protein
mg/L
2
PCT
Procalcitonin
ng/ml