Installation Instructions

Error! Use the Home tab to apply 标题 1 to the text that you
want to appear here.
FA50 Operation Manual 77
Appendix A. Specifications
A.1 Performance Indicators
A.1.1 Repeatability
The repeatability of the analyzer shall comply with one of the
following requirements:
(1)QC card: repeatedly test the reflection ratio of the quality control
(QC) card, the coefficient of variation (CV) of the test results
should not be more than 1%.
(2)Test card: repeatedly test the card of one item, the coefficient of
variation (CV) of the test results should not be greater than
10%.
A.1.2 Accuracy
Test the reference material (calibrator) with a fixed value, the bias
(B) of its average value and marked value should not be greater
than 10%.