Users Manual

11
Control solution is used to check the performance of the kit.
The meter kit should be check:
• When the meter and strip do not work properly.
• When the test result is unusual or inconsistent.
Control solution range is shown as follow:
Please check your strip vial label or single foil pack for exact range.
Perform a Control Test:
Insert an unused strip
Make sure the code number is correct (Ketone)
Press
“S” (3 secs)
Enter control solution mode (shown CS on screen)
Shake the
control solution well
Discard rst three drops
Put one drop onto a clean
surface
Touch the control solution with strip sample inlet
Test result display
after count down
Compare the result with the control range on test strip vial or
single foil pack
Note:
• Do not reuse the test strip.
• Marked open date on new open control solution.
• Do not use the meter if the control test is out of range.
• If the control test keeps result wrongly, please contact an authorized distributor.
• Control solutions are not included. Please contact an authorized distributor for purchasing.