User's Manual

52
put
lines
Surge
IEC 61000-4-5
±1 kV
line(s) to
line(s)
±2 kV
line(s) to
earth
±1 kV line(s)
to line(s)
±2 kV line(s)
to earth
Mains power
quality should be
that of a typic al
commercial or
hospital
environment.
interruptions
and
voltage
variations
on power
supply
input lines
IEC 61000-4-
11
<5 % UT
(>95 %
dip in
UT)
for 0,5
cycle
40 % U
T
(60 %
dip in
U
T)
for 5
cycles
70 % U
T
(30 %
dip in
UT)
for 25
cycles
<5 % UT
(>95 %
dip in
U
T)
for 5 sec
<5 % UT
(>95 % dip in
UT)
for 0,5 cycle
40 % U
T
(60 % dip in
UT)
for 5 cycles
70 % UT
(30 % dip in
UT)
for 25 cycles
<5 % U
T
(>95 % dip in
UT)
for 5 sec
Mains power
quality should be
that of a typic al
commercial or
hospital
environment. If the
user of the
T10YXXXXXX ]
requires continued
operation during
power
mains
interruptions, it is
recommended that
the T10YXXXXXX
be powered from
an uninterruptible
power supply or a
battery.
Power
frequency
(50/60 Hz)
magnetic field
3 A/m
3 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital