14693e - 31 Mar 03  4- 28  Routine Maintenance 
4.6 System Storage  
This storage procedure has been established to store Millenium™ RO machines. It is not 
recommended that a unit be dormant longer than 24 hours without disinfectant. ROs that 
have been left unused for more than 24 hours risk the possibility of colonizing bacteria 
and exceeding acceptable limits. ROs to be stored should first be disinfected with one of 
the procedures in this chapter. The RO should be stored in a location where it cannot 
freeze or become overly hot. 
CAUTION: 
‘RENALIN®’ products (Renalin®, Minncare™, Peracidin™. etc.) are not 
acceptable disinfectants for storage of systems. Exposure of more than 12 hours to 
RENALIN® will result in irreversible damage to the membranes. 
4.6.1  Formaldehyde Storage Procedure  
WARNING: 
Always wear rubber gloves, an apron and a facemask. If the room has ventilation, 
use it. If not, wear a face respirator with an organic cartridge to remove 
formaldehyde fumes. DO NOT BREATHE FUMES OR ALLOW 
FORMALDEHYDE TO COME IN CONTACT WITH SKIN OR EYES Refer to 
the MSDS supplied by the manufacturer of the formaldehyde/formalin product. 
1.  If formaldehyde is chosen to disinfect the RO, follow the instructions for 
formaldehyde disinfection. 
2.  Do not rinse out after the disinfectant has been circulated in the machine. 
3.  Disconnect the disinfect line, the feed water line, the waste line, power and the 
product line. 
4.  Coil the hoses and power cord, and cover the RO with a large plastic bag to 
protect the RO from dust. 
NOTE: 
All ROs can be safely stored in 0.5 to 4 % formaldehyde for up to one year. 
WARNING: 
After storage, the RO unit must be completely rinsed of formaldehyde, and a 
bacterial culture performed and interpreted before use. Perform AAMI analysis if it 
has been one year or more since last done. 