14693e - 31 Mar 03  2 - 4  Installation Guidelines and Instructions 
valve allowing product flow. When the hose is disconnected, the product flow is 
stopped. 
4.  Make sure that the inlet water valve switch and the RO pump power switch are in 
the OFF position. Plug the power cord into a 115 VAC (230 VAC) grounded 
outlet. Do not use an extension cord or an adapter. 
5.  Turn the water supply to the RO on. 
6.  Turn the operate/flush valve to the flush position. 
7.  Turn the inlet water valve switch on. Water will start to fill the RO. Allow the 
RO to rinse to drain for five minutes after water begins to flow from the drain 
line. The pump pressure gauge should indicate some pressure. 
8.  Turn the RO pump power switch on. 
NOTE: 
The operate/flush valve may need to be rotated part way towards the operate 
position if the pump makes excessive noise. 
9.  Slowly turn the operate/flush valve to the operate position. Pump pressure should 
not exceed 200 PSI. If 200 PSI is exceeded, adjust pump pressure on pump head 
located inside the cabinet. See instructions in Figure 1 below: 