User guide
M
Millenium™ RO System by GE Osmonics 
14693e – 31 Mar 03  1 - 12  General Information 
1.10 Specifications: 
1.10.1  Input Water Requirements 
 Minimum  Maximum 
Input flow Rate  2.0 GPM  N/A 
Temperature (water)  1.7°C (35°F)  32°C (90°F) 
NOTE: 
Operating at lower temperatures may make it impossible for the Millenium™ to 
produce adequate permeate for the dialysis machine. 
pH (operating)  4.0  11.0 
 TF Membranes 
pH w/chloramines present (operating)  4.0        8.5 
 TF Membranes 
pH (cleaning)  2.0  11.0 
 TF Membranes 
Pressure, dynamic  30 PSI  100 PSI 
1.10.2 Flow Rates 
Product Flow based on the following conditions: 
 New Membranes 
 Operating Pressure:  190 PSI 
 Product Pressure:  1 PSI 
  Feed Water Temperature:    77°F (25°C) 
GPM LPM GPD LPD 
.34 1.31 500 1892.5 
.52 1.97 750 2839.1 
Membrane flux may vary ±15% 
Waste Flow:  .84 GPM (3.2 LPM) (
±10%) 
1.10.3 Electrical Requirements 
Motor  1/3 HP 
Voltage  115 VAC / 230 VAC 
Hertz  60 / 50 
Amps  8 / 5 
Phase  single 