Brochure

GE Inspection Technologies
Xray
GEInspectionTechnologies.
Tube Shutter Timer
The equipment has two independent, programmable timers with a non-volatile memory. A total of four timers can be supplied if
required. Depending on the number of timers it is possible to time-control one to two or one to four window shutter elements.
Preselection and setting hours from 0 to 99, minutes from 0 to 59, seconds from 0 to 59
digital setting individually for 1 to 2(4) window shutter elements
Set point value display digital, 6 digits
Actual value display digital, 6 digits
Technology
20 kHz IGBT technology for high voltage circuit and filament current circuit
Emitted interference as per EN 55011A, Class A
Immunity to interference as per IEC 801-2/1991, 801-3/1984, 801-4/1988
Connected Loads and Mechanical Dimensions
Power connection 1/N~ 230 V ± 10%, 50/60 Hz
Power consumption 4150 W at 60 kV, 58 mA
Max. mains current 24 A
Mains fuse 32 A (as per VDE)
Ground terminal 6 mm² Cu
Type of protection as per EN 60529, IP00 (rack mount), IP32 (desk-top)
Cooling water connection 3/4 inch
Cooling water consumption minimum 3.5 l/min, minimum 4.5 bar, maximum 7 bar,
at outlet point pressure-free
Cooling water temperature > dew point < 35° C
Dimensions overall 483 mm x 266 mm x 680 mm (WxHxD), 19” (48 cm) rack mount
Weight approximately 57 kg
We reserve the right to technical modifications without prior notice.
©2004 General Electric Company. All rights reserved.
GEIT-30095GB (5/04)