Brochure

Xray
The basic configuration comprises
- Basic equipment ID 3003 as 19”
chassis with integrated high voltage
generator
- Cooling water checking device
- 1 Water hose (ID 3003 - Tubehous
-
ing), standard length 2 m
- 2 Water hoses (inlet hose, outlet
hose), standard length 5 m each
(maximum 20 m *)
- Computer interface RS 232 C
- 20 mA current loop (with additional
adapter)
- RS 422 / 485 (with additional
adapter)
* Hose lengths of more than 5 m require special
diameters to keep pressure loss low
Extensions
- Tubehousing V 4 with X-ray tubes as
per individual choice and respective
separate specifications
- High voltage cable, standard length
2 m (maximum 20 m)
- Tube adapter plug 60 kV
- Desk-top casing 19”/6 HE
- Work table as per individual choice
- Multiple socket outlet 230 V
- Radiation protection cabinet
- Fail-safe warning lamp “X-RAY ON”
- Tubehousing support
- Water cooling aggregate
Sockets for connection of additional
components
- Serial interface, see separate specifi-
cation
- Interlock as per German standard
- Extended interlock, assigned to shut
-
ters
- Additional fail-safe warning lamp
- External START/STOP
- External EMERGENCY-STOP
- Control contacts for peripherals,
assigned to shutters (maximum load
230 V, 2 A)
- 230 V switched at MAINS ON
(230 V, 4 A)
- Control contacts, assigned to MAINS
ON (maximum load 230 V, 4 A)
- 230 V switched at HIGH VOLTAGE
ON (230 V, 4 A)
- Control contacts, assigned to HIGH
VOLTAGE ON (maximum load
230 V, 4 A)
Technical Data
High Voltage Generator
Maximum output voltage 60 kV direct voltage
Maximum output current 80 mA
Maximum output power 3.5 kW
Shunt and discharge resistance 600 M
± 1% TC 25
Tube Voltage
Preselection and setting digital or quasi-continuous from 2 to 60 kV in steps of 1 kV
Set point value display digital, 2 digits
Actual value display digital, 2 digits
Absolute accuracy ± 2% (minimum ± 1 digit)
Stability ± 0.01% at ± 10% mains voltage fluctuations
Ripple < 1%
Tube Current
Preselection and setting digital or quasi-continuous from 2 to 80 mA in steps of 1 mA
Set point value display digital, 2 digits
Actual value display digital, 2 digits
Absolute accuracy ± 1% (minimum ± 1 digit)
Stability ± 0.01% at ± 10% mains voltage fluctuations