Technical Specifications

543
Menu - Calibrate
Reduced Span The value in percentage of calibration scale to reduce the span.
Interface Transmitter:
Low SG/ High SG
For an interface transmitter: This specific gravity of service is nor-
mally the same as the low-high specific gravity used during the
calibration, but can be changed here if the specific gravity of the
process has changed and is not equal to the low-high values of the
calibration.
Level Transmitter: SG
Service
For a level transmitter: The specific gravity of service is normally
the same as the specific gravity of the calibration but can be
changed here if the specific gravity of the process has changed
and is not equal to the calibration values.
Apply Sends all of the changes made to entries on the tab immediately to
the DLT.
Menu - Calibrate
When you right-click on any of the Calibrate tabs, a context menu appears. The items that
appear depend on the HDLT mode:
Report Setup - Allows you to select a report file and a report printing program.
Report - Generates a report of the HDLT parameters.
Detach Trend - Removes the Trend display from the tabbed dialog and creates a
separate trend display.
Help - Displays the help file at the Monitor tab instructions.