Product Manual

Vscan gateway software
4-32 Vscan – User Manual
GM092207 02
Ensure that the power supply is connected to the
Docking station while upgrading the software.
DO NOT TURN OFF the Vscan unit while upgrading the
software.
4. Press OK.
The software update files are copied and unpacked to the
Vscan microSD card.
5. The Vscan is running a system test.
If the test failed, close the display to turn off the system
and contact service.
If the test passed, the software update is installed. This
may take several minutes.
The following messages are displayed on screen.
When the installation is done, the Vscan is turned off
automatically. Close the display.
6. Open the display to restart the Vscan.
The Vscan is running a system test.
If the test failed, close the display to turn off the system
and contact service.
If the test passed the Vscan is ready for use.
NOTE: The next time the Vscan is docked to the Vscan gateway
software, pressing Delete in the Software page will remove the
installed upgrade from the list.
Vscan unit software update from a microSD card
If the software update is provided on a microSD card, proceed
as follows:
CAUTION
Do not turn off or disconnect power during the update.
Doing so could cause an error requiring the device to be
sent for repair.