Acoustic Output Reporting Tables
Vscan – User Manual 8-3
GM092207 02
Explanation of Footnotes
The mechanical and thermal indices may be replaced by one of
the following footnotes because of the reasons listed:
a: Display of this index is not required for this operating
mode.
b: This probe is not intended for transcranial or neonatal
cephalic uses.
c: This formulation for TIS is less than that for an alternate
formulation in this mode.
If so, the table entries are replaced by a “#”, meaning: no data
are provided for this operating condition since the maximum
reported value is not reported for the reason listed.
If neither an index or a footnote is given, this means that the
index is irrelevant for this transducer/mode combination.
Pulse Pressure Squared Integral / Pulse intensity integral
Peak-Rarefactional Acoustic Pressure / (same)
Attenuated Peak-Rarefactional Acoustic Pressure / (same)
Pulse Repetition Rate / Pulse repetition frequency
Thermal Index / (same)
Bone Thermal Index / (same)
Cranial-Bone Thermal Index / (same)
Soft-Tissue Thermal Index / (same)
Pulse Duration / (same)
-12 dB Output Beam Dimensions / (same)
Distance from the Source to a Specified Point / (same)
Depth for TIB / Depth at which the relevant index is maximum
Break-Point Depth / (same)
Depth for TIS / Depth at which the relevant index is maximum
IEC FDA Meaning—IEC 60601-2-37 / FDA & NEMA UD2, UD3