Product Manual

Safety
7-20 Vscan – User Manual
GM092207 02
Maximum probe temperature
NOTE: Lens temperature measured under following conditions per
IEC 60601-2-37:
1. Thermocouple was placed at the geometric center of the
lens.
2. a: Thermal phantom at 37 °C for non-external probes.
b: Thermal phantom at 33 °C (or 23 °C) for external probes.
(Temperature rise is measured and added to 33 °C if the
phantom is at 23 °C).
NOTE: Thermal phantom made with tissue-mimicking material as
referenced in IEC 60601-2-37.
3. Probe placed upright in contact with above thermal
phantom.
4. Auto-freeze capability is disabled.
5. Lens temperature is monitored for 30 minutes.
Probe
Max Temp
(Simulated use)
Max Temp
(Still air)
G3S 40.0 34.5