Product Manual

Important safety considerations
Vscan – User Manual 7-13
GM092207 02
Voltage dips, short
interruptions and
voltage variations on
power supply input lines
IEC 61000-4-11
< 5% U
T
(>95% dip in U
T
) for
0.5 cycle
40% U
T
(60% dip in U
T
) for
5 cycles
70% U
T
(30% dip in U
T
) for
25 cycles
< 5% U
T
(>95% dip in U
T
) for
5sec.
Compliance for all
test levels.
Controlled shutdown
with return to
pre-disturbance
condition after
operators
intervention.
(Power-on switch)
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
the ultrasound unit requires
continued operation during
power mains interruptions, it
is recommended that the
ultrasound unit is powered
from an uninterruptible
power supply or a battery.
Power frequency (50/
60 Hz) magnetic field
IEC 61000-4-8
3A/m 3A/m
50 and 60 Hz
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE: U
T
is the a. c. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration – electromagnetic immunity.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic
environment –
guidance