Product Manual

Vscan maintenance
6-6 Vscan – User Manual
GM092207 02
Disinfecting the probe
1. After cleaning, the probe and cable may be wiped with a
tissue sprayed with a recommended germicide.
WARNING
Use additional precautions (e.g. gloves and gown) when
decontaminating an infected probe.
The probe should not be exposed to the germicide longer than
specified to achieve the desired effect.
DO NOT soak or saturate the probe with solutions containing
alcohol, bleach, ammonium chloride compounds or hydrogen
peroxide.
WARNING
CREUTZFELD-JACOB DISEASE
Neurological use on patients with this disease must be
avoided. If the Vscan becomes contaminated, there is no
adequate disinfecting means.