Operating Manual

151
150
For desktop scanners the cassette can be opened, as
shown in figure 6-16. CR plates can be exposed to
subdued light (< 10 lux: a candle creates 5 lux) for
one minute with acceptable consequences for the
image quality. The effect strongly depends on the
type of the light source, e.g. tube light causes most
damage to the latent image. The scanned image is
ultimately or instantly made visible on a high-reso-
lution monitor (computer screen) of the worksta-
tion, see figure 33-16.
Depending on the line distance selected, typically
50 or 100 microns, the plate speed (lengthwise
progress) is 5 to 10 mm per second. This is similar
to the speed of digitisation of a traditional film. In all scanners, the latent image on the plate
is not only read but also subsequently erased (reset) which takes about one minute, and the-
refore the CR plate is almost immediately available for the next exposure.
CR cassettes
Bare CR plates are nearly as pliable as film. They can be packed in paper or vinyl
cassettes either with or without lead screens. These packages are still pliable.
Technically the plates can be used many times (up to 1000 x), provided they are
handled with utmost care while their surface despite a protective coating is very
sensitive to touching and dirt. A single scratch can make the plate unsuitable for
further use. Rigid cassettes developed especially for the NDT-market have built-in intensify-
ing lead screens at the source side, and a second lead screen at the back to absorb radiation
caused by backscatter. These multi-layer cassettes are not flexible anymore but can be re-
used more often than the flexible cassettes (even a few 1000 times).
Figure 7-16 shows a cross-section of the CR imaging plate in a rigid cassette. The steel and
magnetic plates ensure that the various active layers are evenly and closely pressed
together. For low energy exposures clip-type cassettes exist (to replace the steel- and
magnetic plate) which also ensure intimate contact between the layers. The steel- and
magnet plate would otherwise absorb the low energy radiation.
Dynamic range – Exposure latitude
CR plates have an extremely wide dynamic range (exposure latitude). In practice the
phosphor crystals on a CR plate cannot be saturated and react almost linearly to incident
radiation, while in a conventional film the silver-halide crystals react exponentially, see
the graph in figure 8-16.
As a result the dynamic
range of a CR plate is much
wider than for conventional
film, which makes exposure
times less critical, reducing
re-shoots (re-takes), and
allows various material
thicknesses to be examined
at the same time. The wide
dynamic range can also be
useful in case of under-expo-
sure, this can be compensated
for by a more sensitive read-
out scan or image adjustment
at the work station.
This wide range is illustrated
in figure 9-16. The images
have been obtained from a
stepwedge from 5 up to 25 mm
thickness, in steps of 1mm.
The digitised image of the
film shows only a portion of
the step wedge thicknesses,
the logarithmic CR image
shows all steps proportionally.
The matching analogue records,
at right handside of this figure,
confirm this behaviour.
Furthermore, those sensitivi-
ty (speed) of CR is five to ten
times higher as well, compare
point A and B at a density of
2, see also figure 27-16.
Fig . 6-16. Opened CR cassette
CR plate
Fig. 7-16. Structure of a CR cassette with storage phosphor
Plastic front of cassette
Lead intensifying foil 250
μ
m
Magnetic sheet
Storage phosphor plate
Lead screen 150 μ
m
Steel sheet
Plastic back of cassette
Fig. 8-16. Density/intensity versus dose for film and digital methods
Fig. 9-16. Dynamic range of digitised film versus CR plate. (courtesy BAM Berlin)
Density film
Relative dose
A
B
Film
CR plate
DR panel
Intensity digital methods
Useful
dynamic
range of film
Useful dynamic range of CR/DR
Imaging plate
Imaging plate
Digitised film
Digitised film
Density
Density