Owner manual

InnovOx On-Line Analyzer Details
Specifications
Operating Specifications
1
Analysis Modes NPOC, TOC (TC-IC), TC, IC
Dynamic TOC Range Up to 50,000 ppm TOC
TOC Limit of Detection (LOD) 0.05 ppm NPOC
TOC Accuracy** ±3%
TOC Precision** RSD ≤ 1.5%
TOC Linearity*** R
2
≥ 0.995, measured as NPOC
Analysis Time 2.6 to 8.3 min in online mode
TOC Calibration Stability Up to 6 months
Particle Diameters in Sample ≤ 200 µm diameter without filtration,
> 200 µm with optional Filter Panel
Ambient Temperature Range 10–40 °C (50–104 °F)
Maximum Relative Humidity Up to 95%, non-condensing
Sample Temperature Range 10–60 °C (50–140 °F)
Sample Volume 0.6 ml to 3 ml
Sample Inlet Pressure 125 psig maximum with provided control valve
0.5 psig minimum without optional sample pump
Drain Gravity drain
Analyzer Specifications
Outputs External USB port (1), internal USB ports (2), Ethernet (1), Binary End-of-Analysis Output (1),
Isolated 4-20 mA Analog Outputs (5)
Display Color, touch-sensitive LCD
Power 100–240 ±10% VAC, 280 W, 50/60 Hz
Dimensions Analyzer: H: 92.7 cm (36.5 in); W: 64.6 cm (25.4 in); D: 38.7 cm (15.3 in)
Weight Analyzer: 36.3 kg (80 lb)
Safety Certifications CE, ETL listed. Conforms to UL Std. 61010-1. Certified to CSA C22.2 No. 61010-1.
IP Rating IP56 (optional)
Consumables Maintenance Frequency Estimated Maintenance Time
Acid Reagent As needed 5 minutes
Oxidizer Reagent As needed 5 minutes
Reagent Grade Water As needed 5 minutes
Pumps 12 months 15 minutes
Sample Flow Path Tubing 3 months 15 minutes
Valve Seals (2) Up to 3 months, depending on sample matrix 15 minutes
Sample Pump Tubing (optional) 3 months 15 minutes
0
200
400
600
800
1000
1200
1400
1600
Tim e
Total Carbon (ppm)
Accessories and Options
Air Purifier
The Air Purifier removes CO
2
from the air so the Analyzer can
use house air as the compressed gas source.
Filter Panel
This accessory filters one sample
stream, and it is automatically
cleaned by back flushing with
compressed air. The Filter Panel is
adjustable for various particulate
levels.
Fail-Safe Accessory
Sensors detect when the flow rate of each sample stream
stops or consumables need replenishing.
IP56
The IP56 configuration protects the Analyzer from water
spray and dust that would interfere with its operation.
NEMA 4X
The NEMA 4X enclosure protects the instrument from
corrosive environments.
Other Accessories
Other accessories include a mounting stand, a sample
peristaltic pump, and multi-stream (2, 5) configurations.