User Manual

©2012, General Electric Company. All rights reserved.
Europe/Middle East/Africa
GE Analytical Instruments
Unit 3, Mercury Way
Urmston, Manchester
UK M41 7LY
T +44 (0) 161 864 6800
F +44 (0) 161 864 6829
geai.europe@ge.com
The Americas
GE Analytical Instruments
6060 Spine Road
Boulder, CO 80301-3687 USA
T +1 800 255 6964
T +1 303 444 2009
F +1 303 527 1797
geai@ge.com
Asia Pacic
GE Analytical Instruments
7/F, Building 2, No. 5 Hua Tuo Rd
ZhangJiang Hi-Tech Park, Pudong
Shanghai, China 201203
T +(8621) 38777775
F +(8621) 38777469
geai.asia@ge.com
www.geinstruments.com
ISO
9001:2008
QUALITY
MANAGEMENT
SYSTEM
*Trademark of General Electric Company; may be registered in one or more countries.
**Stated analytical performance is achievable under controlled conditions that minimize operator and standards errors.
The Sievers 500 RL TOC Analyzers are protected by one or more of the following US and foreign patents: US 8,003,048; US 5,976,468; US 5,837,203; US 5,443,991; EP
0897530; FR 0897530; GB 0897530; DE 69702516.0-08. Other patents pending.
For more information, visit www.geinstruments.com. Find a sales partner near you through the “Contact Us” Section.
System Specications**
Total Organic Carbon
Range 0.03 to 2,500 ppb as C
Accuracy
± 0.1 ppb or ± 5% of measurement, whichever is greater
Precision < 1% RSD or 0.03 ppb as C, whichever is greater
pH Range 5.5 to 8.0 in order to meet accuracy specication
Maximum Sample Conductivity 25 µS/cm at neutral pH
Analysis Modes On-Line, On-Line Averaged, On-Line Timed
Analysis Time Continuous mode: 6 minutes
Average and timed modes: 0.5, 1, 4, 8, or 24 hours
External Flow Rate Minimum 50 mL/min
Sample Temperature 1° C to 95° C (34° F to 203° F)
Sample Pressure Up to 100 psig
Interferences Insensitive to organic heteroatoms and measures
organic nitrogen
Calibration Stability Typically stable for 12 months
Display Readout 3 signicant digits
Instrument
Power 100–240 ±10% VAC, 50 W, 50/60 Hz
Normal Operating Environment Intended for indoor use only
Ambient Temperature 10° C to 40° C (50° F to 104° F)
Maximum Relative Humidity Up to 95%, noncondensing
Maximum Altitude 2,300 m (7,546 ft)
Inputs One isolated binary input
Outputs Three isolated 4-20 mA outputs, one Serial (RS-232), one USB
port, one parallel printer port, four alarm outputs, and one
Ethernet port
Installation/Overvoltage Category II (protects against transients present in Category II power)
Safety Certications ETL, CE
Pollution Degree 2 (normally only non-conductive pollution)
Display Backlit Quarter-VGA touchscreen display
Dimensions H: 41.9 cm (16.5 in); W: 48.3 cm (19 in); D: 27.4 cm (10.8 in)
Weight 16.9 kg (37.2 lb)
IP Rating Environmental enclosure: IP 45
Consumables
UV Lamp 6 months
Pump Tubing (with pump heads) 12 months
DI Water Cartridge Typically 24 months (depends on water quality)
Inline lter Replace as needed (depends on water quality)
DI Water Level Rell As needed (check when replacing other items)
300 00165 Rev E