11-32 ApexPro Telemetry System Revision D
2001989-134
ECG Monitoring: ST Analysis
ST Deviation Alarm
When any individual ST value is beyond the limit, an ST deviation alarm
occurs. It is considered a parameter alarm, and the default alarm level is
warning. This can be modified in the parameter alarm level setup.
When the ST program is turned on, or a relearn is done with ST on,
the ST deviation values are set for all leads of ST.
The current ST value is determined in all eight leads.
The ST value in the ECG parameter window flashes to indicate an
alarm.
ST limits can also be adjusted individually in the patient’s Alarm
Control tab.