ECG Monitoring: ECG Tab Sheet
No Arrhythmia Detection with Patient Monitors at 7015 Software Level
If an ApexPro telemetry system patient is admitted to a patient monitor
at the 7015 software level (ECG source is telemetry, not the monitor), the
following scenario may occur when monitoring in Combo or Rover/Combo
monitoring modes:
Since the 7015 software level does not support arrhythmia
processing, arrhythmia detection for the telemetry patient is reduced
from full arrhythmia detection to no arrhythmia detection
(arrhythmia OFF). This occurs because the software is designed to
take on the attributes of the bedside monitor when in Combo or
Rover Combo monitoring modes.
CAUTION
Under these conditions, arrhythmia detection is OFF.
There is NO INDICATION of this at the bedside monitor,
central station or clinical information center.
If the patient is later discharged from the monitor, and monitoring
continues from telemetry, the message ARR OFF will then appear at
the central station or clinical information center. Arrhythmia
monitoring remains OFF.
NOTE
Solar 7000 monitors, Solar 8000 monitors, Dash monitors, and Eagle
monitors may include the 7015 software level.