Revision D ApexPro Telemetry System 11-23
2001989-134
ECG Monitoring: ECG Tab Sheet
Single Lead ECG Telemetry Data
NOTE
When acquiring Single Lead ECG data using a 5-leadwire or 6-
leadwire cable, it is NOT necessary to connect the V leads or the
right leg lead to the transmitter or to the patient.
The following constraints apply when using Single Lead ECG telemetry
data.
Function Single Lead Constraints
change the displayed
lead
The factory default display lead is lead II.
Contact your local service representative to change the
default displayed lead.
Display Lead appears to be selectable at the CIC Pro.
However, your selection is temporary and will revert back
to the ApexPro transmitter’s default displayed lead.
NOTE
When the clinical situation dictates
monitoring a lead other than the default
lead, you can move the leads and/or
electrodes to view a different lead. Be aware
that the label on the display and on the
printout will show the default lead label.
lead analysis
Multi-Lead analysis may appear to be selectable at the CIC
Pro. However, your selection is temporary and will revert
back to the Single Lead analysis mode.
select a V lead
V leads may appear to be selectable at the CIC Pro.
However, your selection does not change the ApexPro
transmitter’s default displayed lead.
select displayed leads
from a single viewer
Leads other than the default displayed lead may appear to
be selectable at the CIC Pro. However, your selection is
temporary and will revert back to the ApexPro transmitter’s
default displayed lead.
select graph waveforms
Leads other than the default displayed lead may appear to
be selectable at the CIC Pro. However, you must select the
ApexPro transmitter’s default displayed lead to obtain a
graph of the waveform.