Operator's Manual Part 2

Table Of Contents
11-20 ApexPro Telemetry System Revision D
2001989-134
ECG Monitoring: ECG Tab Sheet
For successful monitoring of pacemaker patients follow these
suggestions:
Use recommended electrode placement.
Brady, Pause, and Low Heart Rate are additional alarms available
for use when monitoring pacemaker patients.
Problems you may experience are:
heart rate double counting;
inaccurate alarms for low heart rate or asystole;
pacemaker spikes not recognized by the software.
Possible solutions to above problems are:
relearn arrhythmia;
try an alternate electrode placement;
try single-lead analysis;
try switching to the other pace detection mode.
Pacemaker mode:
In most cases, Pace 2 mode will effectively monitor a pacemaker
patient. However, if you are experiencing problems, select the Pace 1
mode as an option, and observe all cautions as described for the
Pace 1 mode of operation.
Multi-Vector Pace Detection (ApexPro CH only)
NOTE
The ApexPro CH transmitter is not for sale outside of the U.S. and
Canada.
The ApexPro CH transmitter uses multi-vector pace detection. Here are
some additional guidelines for successful monitoring pacemaker patients
when using the ApexPro CH transmitter.
When using the 5 or 6-leadwire patient cable with all the electrodes
attached, pace detection occurs on two ECG leads simultaneously.
The default leads used for detection are II and V. If these leads are
not available, multi-vector pace detection switches to available leads.
Pace detection switches to single-lead when using a 3-leadwire
patient cable.
For more information, refer to the Pacemaker Troubleshooting section in
this chapter.