Revision D ApexPro Telemetry System 10-7
2001989-134
Patient Data: Alarm Histories
View Older Event
Click on the View Older Event button to view the next older event
waveforms, if any. If there are no events older than the highlighted
event, this button is dimmed and does not function.
Delete Event
Click on the Delete Event button to delete the event that is currently
highlighted in the event directory. This cannot be undone.
If more than one event is highlighted (using Shift or Ctrl and the mouse),
a message appears when you press the Delete Event button, asking if you
are sure you wish to delete the events. Click on Ok to continue the delete.
Learn As
The Learn As button is inactive for telemetry patients and appears
dimmed.
Relearn
Select the Relearn button to initiate a relearn procedure. A relearn will
clear the patient’s waveforms and templates, and learn the patient’s
current rhythm as the dominant, “normal” rhythm.
Relearn is useful when you have changed a patient’s electrodes or
leadwires, or when it appears that false alarm calls are being made by
the arrhythmia program. The relearn takes only a few seconds.
Caliper
Select the Caliper button to perform caliper measurements on the ECG
waveform. When selected, the Caliper window opens displaying default
ECG waveform and Calipers which allow you to measure time intervals
for the following waveform measurements: PR, QRS, QT, R-R, and ST.