Operator's Manual Part 2

Table Of Contents
8-10 ApexPro Telemetry System Revision D
2001989-134
Alarm Control: Alarm Control Tab
Alarm Off Reason
Selecting from a list of reasons, you can define why alarms have been
turned off. This text is displayed in addition to the ALARMS OFF
message in the patient’s window at the CIC Pro.
The following are the options for alarms off reasons.
Selecting OFF turns all alarms off until you turn them back on.
Selecting any reason establishes an alarm pause state for 5 minutes
in the presence of a valid waveform. After 5 minutes, alarms will
reactivate if the patient is within range of the antenna system for 15
seconds or longer. If the patient remains out of antenna range, the
alarm pause state will continue until the patient re-enters antenna
range for 15 seconds or longer.
NOTE
The patient must be in antenna range for the alarm pause state
to cease.
WARNING
Alarms do not sound, alarm histories are not stored, and
alarm graphs do not print during an alarms off with
reason condition.
Alarms Off Reason Text Displayed At CIC Pro Text Printed On Graph
On (no text) (no text)
Off ALARMS OFF OFF
X-ray XRAY X-RAY
Shower SHOWER SHOWER
Surgery SURGERY SURGERY
Physical therapy P.T. PHYSICAL THERAPY
Cardiac rehab CAR REHAB CARDIAC REHAB
GI Lab GI LAB GI LAB
Occupational therapy O.T. OCCUPATIONAL THERAPY
Off unit OFF UNIT OFF UNT
Cath Lab CATH LAB CARDIAC CATH LAB