D-28 ApexPro Telemetry System Revision D
2001989-134
ApexPro FH Transceiver: Compliance
The ApexPro FH transceiver is intended for use in the electromagnetic
environment specified in the table below. The customer or user of the
device should ensure that is used in such an environment
Table 5. Electromagnetic Immunity for all equipment and systems.
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete, or ceramic
tile. If floors are covered with
synthetic material, then the relative humidity
should be at least 30%.
Electrical Fast Transient/Burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/output
lines
N/A The ApexPro FH transceiver runs only on
battery power.
Surge
IEC 61000-4-5
± 1 kV differential mode
± 2 kV common mode
N/A The ApexPro FH transceiver runs only on
battery power.
Voltage Dips, Short
Interruptions, and Voltage
Variations on Power Supply
Input Lines
IEC 61000-4-11
<5% U
T
a
(>95% dip in U
T
) for 0.5
cycle
N/A The ApexPro FH transceiver runs only on
battery power.
40% U
T
(60% dip in U
T
) for 5
cycles
70% U
T
(30% dip in U
T
) for 25
cycles
<5% U
T
(>95% dip in U
T
) for 5
seconds
Power Frequency
(40/60 Hz) Magnetic Fields
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital
environment.
a. U
T
is the AC mains voltage prior to the application of the test level.