ApexPro FH Transceiver: Maintenance
Cleaning the ECG Leadsets
The transceiver ECG Leadsets are manufactured by Affinity Medical.
Contact your technical support representative for additional leadsets.
WARNINGS
ECG lead wires must be dressed and secured to the
patient to prevent the possibility of them encircling
the patient’s neck and causing strangulation.
Do not use leadsets which exhibit signs of wear or
damage such as cracking or degradation of the
connectors or cable insulation.
Do not sterilize using steam or gamma radiation.
Damage to the leadsets will result.
CAUTIONS
To increase the life of the leadsets, do not pull on the
leadsets to disconnect. Pull gently by grasping the
connectors.
Do not immerse the leadsets in water or other liquid
to clean. Immersion may cause damage to the
leadsets.
Repeated exposure to EtO sterilization will shorten
the effective life of the leadset. The leadsets should be
sterilized only when indicated by specific patient or
hospital requirements.
Make sure that the wires are not twisted around each
other; since, this can interfere with transmission and
produce noise.
Make sure that the lead wires are not inadvertently
pinched in the bed rails. This may cut the insulation
or break the leadset.