ApexPro FH Transceiver: The Basics
Programming Transceivers
Before a transceiver can be used with the ApexPro FH Telemetry
System, it must first be programmed with a Network Key and TTX ID to
match the corresponding CIC Pro. Each ApexPro FH transceiver TTX ID
number must be entered at the corresponding CIC Pro. To enter the TTX
ID on the CIC Pro, refer to the CIC Pro Clinical Information Center
Operator’s Manual.
WARNING
When connected to a non-patient external device, such as a PC
for programming or Diagnostics, the ApexPro FH transceiver
must be removed from the patient. The I/O connector must be
covered, or used with an approved patient accessory, whenever
the ApexPro FH transceiver is connected to the patient.
Failure to keep the port covered when in use could lead to
excessive voltages and currents being applied to the
patient, resulting in cardiac arrest.
The transceiver TTX ID is located on the back label of the transceiver.
Transceiver TTX ID
Deactivate Mode
If two transceivers on the network have the same TTX ID, then the
second transceiver to be powered on will enter Deactivate Mode. In
Deactivate Mode, all LEDs on the front panel will illuminate and flash in
a rotating pattern and the device will not be operational. If the
transceiver enters this condition, contact your system administrator
immediately.