Revision D ApexPro Telemetry System A-5
2001989-134
Message Glossary: Message Glossary
Graph Messages
The following messages appear in the patient’s bed window on the
Clinical Information Center display. These relate to running graphs at
the printer.
OFF NETWORK In the Combo mode, telemetry patient data is
provided from the bedside monitor to the
Clinical Information Center. If the bedside
monitor is removed from the network or
becomes otherwise unavailable, the OFF
NETWORK alarm is generated. The Clinical
Information Center will then display patient
data directly from the telemetry bed along with
this message in that Clinical Information
Center bed window. Should the bedside monitor
reappear on the network, the Combo monitoring
application will automatically resume and the
alarm will be cleared.
This alarm is also generated if the receiver
system is removed from the network or becomes
otherwise unavailable. The bedside monitor
does not sound alarms, but the alarms must be
turned back on at the bedside monitor if they
were previously paused or off.
GRAPH ALARM An alarm graph was initiated and is running.
GRAPH MANUAL A manual graph was requested and is
running.
GRAPH TTX The Graph button on the transmitter was
pressed and a 20-second graph strip is
running.
PRINTING A non-real time graph is currently being
printed.
SAVING An alarm or a manual graph has been
requested but the writer is in use, the writer
door is open, or the writer is out of paper. The
request is saved and will run as soon as the
writer is available.