Operator's Manual Part 2

Table Of Contents
Revision D ApexPro Telemetry System 8-5
2001989-134
Alarm Control: Alarm Control Tab
Alarm Control Tab
Parameter and arrhythmia alarm levels are, from most critical to least
critical, Crisis, Warning, Advisory, and Message.
Each alarm level elicits a specific response from the CIC Pro. The four
alarm levels are configurable. This means that an alarm can be moved
from one category to another if the default is not satisfactory to your
situation. This is done in the Alarm Control tab sheet for the patient
whose alarm settings you wish to adjust.
NOTE
ASYSTOLE and VFIB/VTAC alarms cannot be moved from the
Crisis level.
Accessing the Alarm Control Tab
1. Use the mouse to click in the window of the telemetry patient whose
alarm settings you wish to view or change.
2. The CIC Pro display rearranges as the single patient viewer opens at
the bottom of the display. Click on the patient’s Alarm Control tab to
bring the tab sheet to the front.
The telemetry patient Alarm Control tab sheet is divided into three
sections: Parameter Limits and Alarm Levels, Arrhythmia Alarm Levels,
and Alarms On/Off. The options on this tab sheet are detailed on the
following pages.
421C