Revision D ApexPro Telemetry System 12-13
2001989-134
SpO2 Monitoring: Troubleshooting
NO DATA
The oximeter is still connected, but no valid data is being transmitted to
the receiver. Check that the patient is within antenna range. If the
problem persists, contact GE Medical Systems Information Technologies
Service.
SPO2 PROBE OFF
The probe is off the patient. Check the probe.
The saturation value is X
The oximeter is still connected, but no valid data is being received at the
receiver. The patient may be in an area of poor antenna reception, where
some, but not all data is being transmitted.
This message remains on the CIC Pro for three minutes. If no data is
detected after three minutes, the message changes to NO DATA.