SpO2 Monitoring: Introduction
SpO2 Probe Safety
Be sure to read all literature accompanying probes for specific safety
information. Be aware of the following safety precautions when using
SpO2 probes.
WARNINGS
DATA VALIDITY—Do not expose probe detector to
strong ambient light while monitoring a patient. A poor
signal may result.
DATA VALIDITY—Do not allow tape to block the probe
light detector.
PATIENT SAFETY—Prolonged monitoring may require
changing the probe site periodically. Move the probe if
there is any sign of skin irritation or impaired
circulation. Change the probe site AT LEAST every four
hours to prevent ischemic skin necrosis. If required,
reduce the application periods to HALF the times
recommended above.
PATIENT SAFETY—If a probe is damaged in any way,
discontinue use immediately.
CAUTION
Use only Nonin SpO2 probes with the Apex Oximeter and
Xpod Oximeter. The reliability of SpO2 data obtained
with any other probe has not been verified.