Revision D ApexPro Telemetry System 5-9
2001989-134
Maintenance: Cleaning
More Intensive Disinfecting or Sterilization
CAUTION
The decision to sterilize must be made per your
institution’s requirements with an awareness of the
effect on the integrity of the transmitter and leadwire.
Cleaning the Oximeter and Accutracker DX
To clean the Apex Oximeter, the Nonin Xpod Oximeter, or the
Accutracker DX blood pressure monitor, follow the instructions in
“General Cleaning/Disinfecting” on page 5-5.
CAUTION
Do not autoclave or steam-clean the equipment.
The decision to sterilize must be made per your
institution’s requirements with an awareness of the
effect on the integrity of the transmitter and
leadwire.
Cleaning the Power Unit
Follow these precautions when cleaning the antenna power unit,
connected to the antenna system.
WARNING
When cleaning the power unit, use a cloth dampened
with cleaning alcohol on the outside of the enclosure only.
Do not immerse the product in water or a safety hazard
could arise during use.
Periodic leakage current testing should be done on the combined power
supply and end-use system on a yearly basis when used in a hospital
environment where such test equipment is commonly available.