5-6 ApexPro Telemetry System Revision D
2001989-134
Maintenance: Cleaning
Never use these cleaning agents:
abrasive cleaners or solvents of any kind,
acetone,
ketone,
quaternary ammonium solutions
alcohol-based cleaning agents, or
Betadine
CAUTION
Failure to follow these rules may melt, distort, or dull the
finish of the case, blur lettering on the labels, or cause
equipment failures.
Cleaning the Transmitters
Importance of Proper Cleaning
Some disinfecting solutions can be conductive and cause anomalies in the
performance of the transmitter. Improper cleaning methods can result
in:
appearance of waveforms when the transmitter is not connected to a
patient, causing false alarms,
degradation of overall system performance,
total transmitter failure,
replacement of leadwires and/or transmitter.
Additionally, improper cleaning methods could result in:
melting, dulling, or distorting the case,
blurred lettering on the labels.
NOTE
Cleaning products known to cause the types of problems listed above
include, but are not limited to, Sani-Cloth® wipes and Ascepti®
wipes; these should be avoided. Products that contain active
ingredients and solutions similar to these products should also be
avoided.